Tel: 4008-266-720
E-mail: info@sihanmed.com
Appearance inspection: Check the appearance of the host is intact, and there is no obvious sign of external damage.
2. Fault phenomenon: The host is connected to AC power, and the error ER400 is reported when it is turned on.
3. Disassembly diagnosis: It is judged to be a power module problem. There are damaged components, multiple chips are burned, and internal signals are disordered.
4. Repair plan: ① Repair the power board ③ Calibrate the output voltage and strengthen the internal firmware. (No accessories for bipolar detection.