Force EZ™ Electrosurgical Generator with Instant Response™ Technology;
Instant Response technology provides surgeons with improved performance or lower power settings, minimizing the risk of tissue damage and neuromuscular stimulation, and decreasing the need to "turn up the generator."
Output Waveforms
Bipolar | 470 kHz sinusoid |
Monopolar Cut | |
Pure | 393 kHz sinusoid |
Blend | 393 kHz bursts ofsinusoid, recurring at 27 kHzintervals. 50% duty cycle envelope |
Monopolar Coag | |
Desiccation | |
Low 1 | 240+40 kHz sinusoidrecurring at 39 kHz. 8% dutycycle |
Low 2 | 393 kHz sinusoid |
Low 3 | 393 kHz sinusoid |
Fulguration | |
High 1 | 470 ± 40 kHz dampedsinusoidal bursts with a |
repetition frequency of 57 kHz | |
High 2 | 470 ± 40 kHz dampedsinusoidal bursts with a |
repetition frequency of 30 kHz |
Output power changes byless than 15% or 5 watts,whichever is greater, as the line voltage varies from 90-135volts (into a 300 ohm load).
Dimensions and Weight
Width | 40.6 cm (16 in.) |
Depth | 39.5 cm (15.6 in.) |
Height | 12.7 cm (5 in.) |
Weight | < 6.8 kg (< 15 lbs) |