Stryker Camera 1288-210-105 Repair Case

Appearance inspection: The camera has normal wear and tear, and the cable has slightly aged.

Fault detection: Test on the machine, shake and pull the cable at the camera end, and the screen will have stripe interference, flickering, and colorful screen. Connect the mirror, the image is slightly to the right, which does not affect the use and can be left untreated for the time being.

Fault cause: The internal signal line of the cable has poor contact, resulting in stripe interference, flickering, and colorful screen.

Repair plan: Replace with a new cable